Optical testing the compliance with regulatory standards of urea concentration in AdBlue®

Carlo Anelli, Vanessa Pellicorio, Valentina Bello, Sabina Merlo
Abstract:
This paper presents an absorption-based optical approach to check the compliance of urea concentration in AdBlue® with regulatory standards. The implemented low-cost and compact measuring system exploits two LEDs that emit in the SWIR spectral band, a glass capillary, and an amplified photodiode to detect the intensity of the radiation transmitted through the fluid under test. The functionality of the system has been demonstrated on urea–water solutions with urea concentrations from 0 up to 525 mg/mL. Thanks to the attained calibration curve, we retrieved the urea concentration of different AdBlue® samples, found in agreement with HPLC results. The proposed optical method allows to easily verify if the estimated urea concentration falls within the required range for AdBlue®.
Download:
IMEKO-TC2-2025-020.pdf
DOI:
10.21014/tc2-2025.020
Event details
IMEKO TC:
TC2
Event name:
IMEKO TC2 PhotoMet 2025
Title:

2025 IMEKO TC2 International Symposium on Modern Photonic Metrology

PhotoMet 2025 - Shaping the Future of Photonic Metrology

Place:
Modena, ITALY
Time:
01 September 2025 - 03 September 2025