Abstract:
Mass determination by use of mass comparators is influenced by different parameters including the conditions of ambient air. Having a look at the temperature of weights and the gravitational field gradients, the relatively small impact of these parameters on mass determination can be easily determined on the base of reproducible measurement results. Of metrological interest is the air density determination and control during a measurement series that could be used to improve the quality of measurement. The influence of temperature change as a function of time and its control simultaneously during the weighting process is discussed. The implementation and use of small climate chambers around the mass comparators to smooth out the temperature and relative moisture variations is described and discussed.